UF75mplus
Standard delivery Universal oven Um
This universal oven made of high-quality hygienic, easy to clean stainless steel leaves nothing to be desired in terms of ventilation technology, control technology, overtemperature protection and perfectly tuned heating technology.
Temperature
Working temperature range |
at least 5 (UN/UNplus/UNm/UNmplus) or 10 (UF/UFplus/UFm/UFmplus) above ambient temperature to +300 °C |
Setting accuracy temperature |
up to 99.9 °C: 0.1 / from 100 °C: 0.5 |
Setting temperature range |
+20 to +300 °C |
Temperature sensor |
2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error |
Control technology
ControlCOCKPIT |
TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays. |
Language setting |
German, English, Spanish, French, Polish, Czech, Hungarian |
Timer |
Digital backwards counter with target time setting, adjustable from 1 minute to 99 days |
Function HeatBALANCE |
adapting the distribution of the heating performance of the upper and lower heating circuit from -50 % to +50 % |
Function SetpointWAIT |
the process time does not start until the set temperature is reached |
Calibration |
three freely selectable temperature values |
adjustable parameters |
temperature (Celsius or Fahrenheit), fan speed, air flap position, programme time, time zones, summertime/wintertime |
Ventilation
Fan |
forced air circulation by quite air turbine, adjustable in 10 % steps for each segment individually |
Fresh air |
Admixture of pre-heated fresh air by electronically adjustable air flap |
Vent |
vent connection with restrictor flap |
Communication
Documentation |
programme stored in case of power failure |
Programming |
AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port |
Safety
Temperature control |
mechanical temperature limiter TB, protection class 1 according to DIN 12880 to switch off the heating approx. 20°C above nominal temperature |
Temperature control |
overtemperature monitor TWW, protection class 3.1 or adjustable temperature limiter TWB, protection class 2, selectable on display |
AutoSAFETY |
additionally integrated over- and undertemperature monitor "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating is switched off in case of overtemperature |
Autodiagnostic system |
for fault analysis |
Alarm |
visual and acoustic |
Standard equipment
Works calibration certificate |
Calibration at +160°C |
Door |
fully insulated stainless steel door with 2-point locking (compression door lock) |
Internals |
2 stainless steel grid(s), electropolished |
Stainless steel interior
Dimensions |
w(A) x h(B) x d(C): 400 x 560 x 330 mm (d less 39 mm for fan) |
Interior |
easy-to-clean interior,made of stainless steel, reinforced by deep drawn ribbing with integrated and protected large-area heating on four sides |
Volume |
74 l |
Max. number of internals |
6 |
Max. loading of chamber |
120 kg |
Max. loading per internal |
20 kg |
Textured stainless steel casing
Dimensions |
w(D) x h(E) x d(F): 585 x 944 x 514 mm (d +56mm door handle) |
Housing |
rear zinc-plated steel |
Electrical data
Voltage
Electrical load |
230 V, 50/60 Hz approx. 2500 W |
Voltage
Electrical load |
115 V, 50/60 Hz approx. 1800 W |
Ambient conditions
Set Up |
The distance between the wall and the rear of the appliance must be at least 15 cm. The clearance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby appliances must not be less than 5 cm. |
Altitude of installation |
max. 2,000 m above sea level |
Ambient temperature |
+5 °C to +40 °C |
Humidity rh |
max. 80 %, non-condensing |
Overvoltage category |
II |
Pollution degree |
2 |
Packing/shipping data
Transport information |
The appliances must be transported upright |
Customs tariff number |
8419 8998 |
Country of origin |
Federal Republic of Germany |
WEEE-Reg.-No. |
DE 66812464 |
Dimensions approx incl. carton |
w x h x d: 730 x 1130 x 670 mm |
Net weight |
approx. 66 kg |
Gross weight carton |
approx. 85 kg |
Standard units are safety-approved and bear the test marks
Intended use as medical device
Memmert is bringing medical devices of class I according to EU Directive 93/42/EEC into circulation until the regulation MDR (EU) 2017/745 comes into force. After the new regulation comes into force, Memmert will bring medical devices of class I into circulation according to MDR.
Type |
Intended use as medical device |
Classification according to MDD |
UNmplus, UFmplus |
The appliance may be used for heating fango, silicate and APS packs for physical therapy and keeping them warm. |
I |
UNm, UFm, INm, IFm |
The appliance may be used for heating fango, silicate and APS packs for physical therapy and keeping them warm. |
I |
INmplus, IFmplus |
The appliance may be used for temperature control of rinsing and infusion solutions and contrast agents. |
I |
IFbw |
The appliance may be used for heating non-sterile covers and cloths. |
I |
Memmert is bringing medical devices of class IIa and IIb into circulation according to MDD 93/42/EEC until 26 May 2024 according to the transitional provisions by the (EU) 2017/745 article 120 (2).
Type |
Intended use as medical device |
Classification according to MDD |
SNplus, SFplus |
The product is intended for the sterilisation of medical devices with dry heat at atmospheric pressure. |
IIb |
SN, SF |
The product is intended for the sterilisation of medical devices with dry heat at atmospheric pressure. |
IIb |
ICOmed |
The appliance is intended for the creation and maintenance of constant environmental conditions for application in the field of in vitro fertilisation (IVF), especially for the incubation of oocytes, spermatozoa and zygotes in special culture dishes for IVF application as well as for gene expression and the biosynthesis of RNA and proteins. |
IIa |
Download as pdf
As part of our medical device consultant training, we qualify you in the consultation, use and instruction of our medical products, taking into account the currently applicable legal requirements with regard to the new Medical Devices Regulation (EU) 2017/745. We explain the requirements for dealers, the geographical scope as well as information on the reporting system and the tasks of a medical device consultant. After successful participation, you will receive a certificate of qualification as a medical device consultant for Memmert medical devices.
Have we piqued your interest? Please contact us at sales@memmert.com for more information.